Advanced RT Objects Interoperability

Name:
Advanced RT Objects Interoperability
ID:
ARTI
State:
7 - Superseded
Detail:
Superseded by Treatment Planning - Plan Content
DocumentVersion:
1.6
Started:
2004
LastUpdate:
2014-02-25
LatestDocument:
ARTI Supplement v1.6 ARTI Spreadsheet v1.4
Overview:
Advanced RT Objects Interoperability
ClinicalImpact:
ARTI Clinical Impact Statement
ProfileProposal:
ARTI Proposal
EntryUpdate:
2020-10-06
MainAuthor:
Bruce Curran
CoAuthor:
Christof Schadt
Reviewer:
Jim Percy
FinalCPs:
FIXME
WorkCPs:
FIXME

Overview

List of actors

Incomplete list:

  • Basic static (No MLC, needs blocks) Produce
  • Basic static (No MLC, needs blocks) Consume
  • RO-ARTI-33 Basic static MLC (like basic static, but with MLC, only 2 C/P) Produce
  • RO-ARTI-34 Basic static MLC (like basic static, but with MLC, only 2 C/P) Consume
  • Motorized Wedge Beam Consume/Produce
  • Hard Wedge Beam Consume/Produce
  • Virtual Wedge Beam Consume/Produce
  • Basic Arc (No MLC, contains blocks, same shape for whole beam (Produce)
  • Basic Arc (No MLC, contains blocks, same shape for whole beam (Consume)
  • RO-ARTI-11 Conformal arc (change of shape ok, no mod) Produce
  • RO-ARTI-12 Conformal arc (change of shape ok, no mod) Consume
  • RO-ARTI-13 Step 'n' shoot / SMLC Produce
  • RO-ARTI-14 Step 'n' shoot / SMLC Consume
  • RO-ARTI-15 Sliding window /DMLC Produce
  • RO-ARTI-16 Sliding window /DMLC Consume
  • RO-ARTI-21 IMAT/VMAT (modulation of beam) Produce
  • RO-ARTI-22 IMAT/VMAT (modulation of beam) Consume
  • RO-ARTI-35 MLC Arc (same MLC shape all over) Produce
  • RO-ARTI-36 MLC Arc (same MLC shape all over) Consume

Test instructions

Planning Instructions

  • Isocenter position is in DICOM coordinates, not IEC
  • Jaw positions are missing a minus-sign, should be X1=-4, X2=6, Y1=-4, Y2=6
  • For Varian machines use dose rate 600 MU/min or lower. 800 and 1000 MU/min are reserved for stereotactic treatments.
  • Note that ARTI does not involve import/export of RT Dose. The only dose involved is the beam dose specification point (located in the RT Plan)

Discussions

Log In