Advanced RT Objects Interoperability
- Name:
- Advanced RT Objects Interoperability
- ID:
- ARTI
- State:
- 7 - Superseded
- Detail:
- Superseded by Treatment Planning - Plan Content
- DocumentVersion:
- 1.6
- Started:
- 2004
- LastUpdate:
- 2014-02-25
- LatestDocument:
- ARTI Supplement v1.6 ARTI Spreadsheet v1.4
- Overview:
- Advanced RT Objects Interoperability
- ClinicalImpact:
- ARTI Clinical Impact Statement
- ProfileProposal:
- ARTI Proposal
- EntryUpdate:
- 2020-10-06
- Reviewer:
- Jim Percy
- FinalCPs:
- WorkCPs:
Overview
List of actors
Incomplete list:
- Basic static (No MLC, needs blocks) Produce
- Basic static (No MLC, needs blocks) Consume
- RO-ARTI-33 Basic static MLC (like basic static, but with MLC, only 2 C/P) Produce
- RO-ARTI-34 Basic static MLC (like basic static, but with MLC, only 2 C/P) Consume
- Motorized Wedge Beam Consume/Produce
- Hard Wedge Beam Consume/Produce
- Virtual Wedge Beam Consume/Produce
- Basic Arc (No MLC, contains blocks, same shape for whole beam (Produce)
- Basic Arc (No MLC, contains blocks, same shape for whole beam (Consume)
- RO-ARTI-11 Conformal arc (change of shape ok, no mod) Produce
- RO-ARTI-12 Conformal arc (change of shape ok, no mod) Consume
- RO-ARTI-13 Step 'n' shoot / SMLC Produce
- RO-ARTI-14 Step 'n' shoot / SMLC Consume
- RO-ARTI-15 Sliding window /DMLC Produce
- RO-ARTI-16 Sliding window /DMLC Consume
- RO-ARTI-21 IMAT/VMAT (modulation of beam) Produce
- RO-ARTI-22 IMAT/VMAT (modulation of beam) Consume
- RO-ARTI-35 MLC Arc (same MLC shape all over) Produce
- RO-ARTI-36 MLC Arc (same MLC shape all over) Consume
Test instructions
- Isocenter position is in DICOM coordinates, not IEC
- Jaw positions are missing a minus-sign, should be X1=-4, X2=6, Y1=-4, Y2=6
- For Varian machines use dose rate 600 MU/min or lower. 800 and 1000 MU/min are reserved for stereotactic treatments.
- Note that ARTI does not involve import/export of RT Dose. The only dose involved is the beam dose specification point (located in the RT Plan)