Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revision Previous revision
connectathon:arti [2018/09/11 11:06]
shadley
connectathon:arti [2018/09/11 11:15] (current)
shadley old revision restored (2016/10/18 11:54)
Line 1: Line 1:
  
-===== IHE Radiation Oncology - TPPC Profile Testing ===== +===== IHE Radiation Oncology - ARTI Profile Testing ===== 
-Treatment Plan Plan Content ​(TPPC) Profile is tested by importing CT image and Structure Set test data in a Producer Actor, creating a treatment plan (and dose) for a supported beam type, and exporting the plan (and dose) to an Archive for retrieval by a test partner (Consumer Actor). ​ Side-by-side comparison of plan displays on the Producer and Consumer Actors is used to verify interoperable exchange of plan information.+The Advanced RT Integration ​(ARTI) Profile is tested by importing CT image and Structure Set test data in a Producer Actor, creating a treatment plan (and dose) for a supported beam type, and exporting the plan (and dose) to an Archive for retrieval by a test partner (Consumer Actor). ​ Side-by-side comparison of plan displays on the Producer and Consumer Actors is used to verify interoperable exchange of plan information.
  
  
Line 9: Line 9:
  
   * Place isocenter as follows: ​   * Place isocenter as follows: ​
-    * **Supine patient ​H&N:** x = -3.2mm, y = 69.2mm, z = 239 mm (DICOM) +    * **Supine patient:** x = -3.2mm, y = 69.2mm, z = 239 mm (DICOM) 
-    * **Prone patient ​Pelvis**: x = -2.7mm, y = 28.4mm, z = 254 mm (DICOM) +    * **Prone patient**: x = -2.7mm, y = 28.4mm, z = 254 mm (DICOM) 
-    * **Brain mets patient**: ​CT isocenter 1: x = 42.3mm, y = -194.3mm, z = -65.0mm; ​ isocenter 2: x = -10.2mm, y = -256.8mm, z = -66.0mm (DICOM) ​    +    * **Brain mets patient**: isocenter 1: x = 42.4mm, y = -194.3mm, z = -65.0mm; ​ isocenter 2: x = -10.2mm, y = -256.8mm, z = -66.0mm (DICOM) ​                          
-    * **Brain mets patient**: MR isocenter 1: x = 36.8mm, y = 17.9mm, z = 32.7mm; ​ isocenter 2: x = -14.7mm, y = -45.7mm, z = 32.0mm (DICOM) ​                          +
   * Specify a total dose of 54 Gy (27 x 2 Gy fractions) to ptv54, unless otherwise specified.   * Specify a total dose of 54 Gy (27 x 2 Gy fractions) to ptv54, unless otherwise specified.
   * Label the plan appropriately:​ e.g., plan_static_mlc   * Label the plan appropriately:​ e.g., plan_static_mlc